Abstract

A direct ion chromatographic method for the determination of chlorite, chlorate, and bromate in the presence of fluoride, chloride, nitrate, nitrite, and bromide in treated drinking waters was described. Separation of target analytes was achieved using an AS19-HC analytical column (250 mm x 4 mm), AG 19- HC guard column (50 mm x 4 mm), and KOH 25 mmol/L as mobile phase. Inorganic analytes were eluted using a flow rate of 1 mL/min. The column temperature was set up and maintained at 300C. The analyte ions were quantified using a suppressed conductivity detector.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.