Abstract

The authors present a novel method for calibrating the diode detectors used in six-port network analyzers without disconnecting the diodes. The method uses either a precision variable attenuator at the six-port input or a power meter connected to the six-port test port. No actual measurement of power is required to calibrate the diode detectors when an attenuator is used. The method is very well suited to in situ calibration of the diode detection circuits for the operation of the six-port with pulsed signals when the average power detected by the power meter drops to very low levels for low duty cycles. In these cases. The precision variable attenuator method is required.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.