Abstract
A compact, high resolution X-ray and gamma ray spectroscopy system has recently been developed which provides excellent energy resolution over a wide range of photon energies. Two detector types have been tested, a cooled silicon PIN photodiode for X-ray energies from 2keV to 30 keV and a CdTe detector for energies from 20 keV to 1 MeV. The CdTe detector can be operated cooled or at room temperature. The heart of the system is a miniature preamplifier developed at AMPTEK which is used in combination with a small thermoelectric cooler to reduce the leakage current of the detectors. Using the cooled Si-PIN detector, 660 eV FWHM was achieved at 5.9 keV and using a CdTe detector 1.3 keV was achieved at 60 keV both with nearly 100% counting efficiency. Good resolution was also achieved at higher energies.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.