Abstract

Objectives To determine the deterioration in the maximal active deflection and flow capabilities of a new-generation flexible ureteroscope (URS). Methods We performed 50 ureterorenoscopies using the same URS: a Karl Storz 11278 AU1 Flex-X, which is capable of a 270° active deflection in the ventral and dorsal directions. Postoperatively, we evaluated the maximal active ventral and dorsal deflection, irrigation flow at 100 cm H 2O, and the number of broken optical fibers. Results The URS was used for a total of 76 hours, 15 minutes (average 95 minutes per procedure). The maximal ventral deflection had deteriorated from 270° to 208° at the last procedure. The maximal dorsal deflection had decreased from 270° to 133°. The irrigation flow at 100 cm H 2O had decreased from 50 mL/min initially to 40 mL/min at the last procedure. The number of broken image fibers had reached six at the 50th procedure. One repair was necessary at the 50th procedure because of a laser perforation of the working channel. Conclusions The need for repair occurs less frequently with the new-generation of URSs, especially when they are used by an experienced endourologist. In the future, improvements in the durability of the flexible URS will still be key for widespread use of the technique and especially to limit damage by the laser to the working channel.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.