Abstract
The efficiency of analyzing nitrogen by optical emission spectroscopy was put to test in a first publication by six German laboratories of the iron and steel industry. Further experiences with this new method accumulated in the meantime will be reported here. By taking into account interelement effects both the residual scatter and BEC (background equivalent concentration) values for alloy steels have clearly been improved. The detection limit was determined at 6 μg/g by using the repeatability standard deviation in the range of the spectral background. Precipitates of titanium nitrides or carbon nitrides lead to high scatter of measured values. The long-term stability of the spectrometric measuring system is largely dependent on the purity of the argon gas, transmission conditions in the spectral apparatus, the properties of the excitation unit, and the quality of the specimen surface. To all these items recommendations for practice are given.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.