Abstract
A new fiber-optic attenuator calibration technique has been developed using linear intensity modulation characteristics of a laser diode. It does not require detector linearity to be known. The uncertainty of attenuation calibration for a typical 10 dB attenuator is 0.020 dB at two standard deviation levels. The resolution of the system is 0.001 dB and the dynamic range is 25 dB in a single step.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.