Abstract
In this work, a new double quantum filter for glutathione (GSH) editing is introduced, combined with the point-resolved spectroscopy localization sequence (PRESS), and demonstrated in vivo. Compared to the conventional double quantum coherence filter, the new filter has two major advantages. First, it eliminates the need for calibration scans for optimizing the signal yield, making it more efficient and convenient for routine use. Second, it removes the influence of water saturation pulses on the GSH yield, further improving its accuracy. With this method, GSH concentrations in the left and right parietal lobes of five healthy volunteers were determined to be 0.91+/-0.16 mM and 0.89+/-0.16 mM, respectively, in agreement with previous studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.