Abstract
We are developing Ta-based superconducting tunnel junction (STJ) X-ray detectors for high-resolution soft X-ray spectroscopy at the synchrotron. For scaling to large detector arrays, we have also built a compact, low-cost, remote-controllable preamplifier with <3 eV electronic noise. Current Ta-STJs attain an energy resolution between 6.5 and 9 eV FWHM for energies up to ∼2 keV, and can be operated at rates up to ∼5,000 counts/s as long as the signals decay with a single exponential time constant.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.