Abstract

A new-designed measurement device for radon and thoron activity concentration is developed based on gas direct measurement to support their in-situ calibration. It consists of a 2000 mm2 Passivated Implanted Planar Silicon (PIPS) detector, a Multi-Channel Analyzer (MCA), a Micro Controller Unit (MCU), and a small electrostatic chamber with a volume of nearly 23 ml. The device records those alpha particles emitted from radon and thoron gas, and the detection efficiency and the crosstalk factor of 218Po/216Po are determined by Monte Carlo simulation. Measurement results have been compared with AlphaGUARD DF2000 in pure radon and thoron environments, respectively. Results show that the measurement results of the devices and the reference monitor agree well with each other, with an average relative deviation of 0.48% for radon gas from about 3300 Bq/m3 to 38 kBq/m3 and -3.25% for thoron gas from about 25 kBq/m3 to 70 kBq/m3. Uncertainty assessment has also been done, and a relative system uncertainty of radon is about 6.8%, while that of thoron is nearly 7.3%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.