Abstract
An in vivo system was developed which allows researchers to determine the intraluminal pressures necessary to burst anastomosed blood vessels. Rat femoral arteries underwent standard microsurgical anastomoses and sham surgeries. After 6 min of leak-free blood flow, the vessels were ligated and cannulated with PE 50 tubing. The catheter was tied into position with 6-0 suture, constructing a water-tight balloon of tissue inclusive of the repair site around the tubing, then attached via a "Y" adapter to a pressure transducer with a digital readout and to a saline filled syringe in an infusion pump. The digital readout from the transducer accurately expressed the pressure on the anastomosis. The highest pressure recorded when the anastomosis ruptured was taken to be the anastomotic burst pressure. Mean burst pressure for 9 suture anastomoses was 326 mm mercury. All vessel burst pressures exceeded rat mean systolic pressure; 90% exceeded rat maximum systolic pressure. All sham surgeries failed to burst, even at pressures exceeding 1,500 mm Hg. The burst pressure test clearly provides researchers with a valid quantitative measure for in vivo comparison of the efficacies of various anastomotic techniques.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.