Abstract

A new bipolar diathermy forceps system was developed to solve the problems of constant, pressure-limited flow rate, and one-sided irrigation. A roller pump, activated synchronously by pressing a foot switch, feeds dripping and flushing solution to the target tissue via the tip at both ends of the forceps. This system is volume-limited. Continuous compression of the foot switch first activates the flushing function, which continues for less than 1 second, during which time bleeding spots can be detected. The flow then changes automatically to the dripping function to suppress tip burning and prevent damage to the surrounding tissues from heat and current leakage. Repeated pressing of the foot switch initiates the jet irrigation function (continuous high flow rates), allowing irrigation of hematomas and removal of excess debris.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.