Abstract
A new Beam Delivery System (BDS) has been studied in the framework of a new proton therapy project, called AMIDERHA. It is characterized by an active scanning system for target irradiation with a pencil beam. The project is based on the use of a Linac with variable final energy and the Robotized Patient Positioning System instead of the traditional gantry. As a consequence, in the active BDS of AMIDERHA a pencil beam scanning system with a relatively long Source to Axis Distance (SAD) can be used. In this contribution, the idea of using a unique new device capable of both horizontal and vertical beam scansion for the AMIDERHA active BDS will be presented and discussed. Furthermore, a preliminary design of that device will be shown, together with the results of simulations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.