Abstract

High-resolution neutron fluence spectra have been measured in the National Accelerator Centre's p(66)/Be(40) neutron therapy beam by the pulsed-beam time-of-flight method. ICRU muscle kerma spectra have been derived from the fluence spectra. Spectral changes resulting from different irradiation conditions have been quantified in terms of the average neutron energy and the fractional low-energy (< 16 MeV) contribution. The changes observed with different thicknesses of polyethylene filtration are consistent with changes in quality parameters determined in biological and microdosimetric experiments. The dosimetry parameters (KtissueA150) N and (Wgas) N calculated for the measured spectra agree with the values recommended in the neutron dosimetry protocol. The shapes of the present fluence spectra differ from previous measurements of p(> 40)/Be spectra. In particular, they differ significantly from the spectrum measured by recoil techniques in an identical neutron therapy unit at the Clatterbridge Hospital, UK. The reasons for the difference are not known.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.