Abstract
Two miniature and compact optoelectronic devices fabricated by means of integration of light emitting diodes have been developed for turbidimetric and nephelometric measurements. These devices are operating according to paired-emitter-detector-diode (PEDD) principle. The detectors have been characterized using bovine serum albumin and Exton protein assay as a model analyte and a model analytical method, respectively. The developed detectors have been adapted for measurements under conditions of flow injection analysis (FIA). Under optimized conditions the turbidimetric flow system offers the range of linear response up to 400mgL−1 with the detection limit at 20mgL−1. The linear range and detection limit found for optimized nephelometric FIA system are 15–500mgL−1 and 8mgL−1, respectively. The PEDD-based FIA systems with the detector operating according to both modes of measurements have been successfully applied for urinalysis offering total protein determination at physiological and pathological levels with high throughput (over 60 injections per hour).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.