Abstract

To compare current practice of cuff size selection for noninvasive blood pressure measurement in a single-centre, tertiary-level neonatal intensive care unit (visual assessment of bladder width/limb length closest to 2/3) with common recommendations for appropriate cuff selection. Visual assessment of the appropriate cuff size ('2/3 rule') for upper arm, forearm and calf in 103 neonates (309 cuff selections) was compared with the following recommendations: (i) Method A - guidelines of the cuff manufacturer, (ii) Method B - cuff width/limb circumference ratio 0.44-0.60 and (iii) Method C - cuff width/limb length ratio closest to 0.66. The upper arm cuff size was correctly chosen in 84% of cases (Method A), 43% (Method B) and 56% (Method C). The forearm cuff size was correctly chosen in 94% of cases (Method A), 68% (Method B) and 54% (Method C). The calf cuff size was correctly chosen in 96% of cases (Method A), 72% (Method B) and 63% (Method C). The accuracy of selecting cuff size by visual assessment is low. Further research on accurate cuff selection for neonates, including at the forearm and calf, is warranted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.