Abstract

A pressure comparison in the negative gauge pressure range was arranged in 2009. The participating laboratories were CMI/Czech Republic, FORCE Technology/Denmark, AS Metrosert/Estonia and MIKES/Finland. Negative gauge pressures are a common range for pressure calibrations although uncertainty requirements are generally not very high.The results from the four participating laboratories suggest that calibrations in the negative gauge pressure range are not as easy as expected. Some of the claimed uncertainties were perhaps too optimistic, and the large variation in the results made it difficult to generate consistent reference values. The agreement of the results at positive gauge pressures on the same transfer standard was much better. Obviously there is a need for further comparisons in the negative gauge pressure range.The transfer standard was a multifunction calibrator Beamex MC5 equipped with an internal pressure module for the range −100 kPa to 104 kPa in the gauge mode. The resolution of the display was 0.001 kPa. The stability of the transfer standard was good.The comparison was registered as EURAMET Project No. 1131 and as the supplementary comparison EURAMET.M.P-S8 in the BIPM key comparison database.Main text.To reach the main text of this paper, click on Final Report.Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by EURAMET, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call