Abstract
The effects of applied voltages and reaction times on negative ion chemical ionization in the quadrupole ion trap are investigated. Mass-selected ejection of undesired reagent ions and selective mass storage of only negative ions are required for practical negative ion chemical ionization. This is achieved by application of rf and dc voltages to the ring electrode to control the mass-to-charge ratios (one polarity) of ions stored, as well as by application of a supplemental rf voltage applied across the endcap electrodes to selectively eject ions of a particular mass-to-charge ratio. Even with careful control of these parameters, negative chemical ionization is not as sensitive as electron ionization and positive chemical ionization because of the lack of thermal electrons in the ion trap. Mass selection of the hydroxide anion as a reagent ion and exclusion of all positive ions provide [M − H] − ions with little or no fragmentation for a wide variety of compounds.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Society for Mass Spectrometry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.