Abstract
Ventricular tachycardia associated with papillary muscle (PM) is often refractory to standard radiofrequency ablation (RFA). The needle-tipped ablation catheter (NT-AC) has been used to treat deep intramyocardial substrates, but its use for PM has not been characterized. Using an ex vivo experimental platform, both 3mm and 6mm NT-AC created larger ablation lesion volumes and depths than open-irrigated ablation catheter did (OI-AC; e.g., 57.12 ± 9.70mm3 and 2.42 ± 0.22mm, respectively; p < 0.01 for all comparisons). Longer NT-AC extension (6mm) resulted in greater ablation lesion volumes and maximum depths (e.g., 333.14 ± 29.13mm3 and 6.46 ± 0.29mm, respectively, compared to the shorter 3mm NT-AC extension, 143.33 ± 12.77mm3, and 4.46 ± 0.14mm; both p < 0.001). There were no steam pops. In conclusion, for PM ablation, the NT-AC was able to achieve ablation lesions that were larger and deeper than with conventional OI-AC. Ablation of PM may be another application for needle-tip ablation. Further studies are warranted to establish long-term safety and efficacy in human studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.