Abstract

Frequency calibration of a 1.3-microm diode laser spectrometer was carried out with a simultaneously recording spectrum of the iodine molecule in the visible region using internal second harmonics from the diode laser itself. An absolute accuracy of 0.002 cm(-1) was achieved with the frequency standard spectrum of the iodine molecule in the visible region and frequency markers from an étalon. Several atomic and molecular transitions in the near-infrared region were remeasured using this new frequency reference.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.