Abstract
We have investigated the application of near-IR reflectance spectroscopy to the determination of motor oil contamination in sandy loam. Although the present work is concerned with a specific case of contamination, we discuss the possibility of applying the method to other organic contaminants and other types of soil. The spectral region considered was 1600–1900 nm, which contains the first overtone of the CH stretch. Using a commercial Fourier transform spectrometer together with cross-validated partial least-squares data analysis, the one-sigma precision for the determination of motor oil in sandy loam was 0.17 wt % (0.13 to 0.26 wt % at the 95% confidence level). The largest contribution to the precision of the determination was sampling error, or inhomogeneity in each sample. Given the precision limit imposed by the sampling error, we found that the performance of the spectrometer could be lowered without affecting the overall precision. In a modeling exercise, adequate performance was obtained with a spectrometer having only seven spectral channels with a spectral resolution of 10 nm and a spectral noise level of 10−3 absorbance units. A design for an inexpensive miniature instrument is presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.