Abstract

A portable fundus imager is essential for emerging telemedicine screening and point-of-care examination of eye diseases. However, existing portable fundus cameras have limited field of view (FOV) and frequently require pupillary dilation. We report here a miniaturized indirect ophthalmoscopy-based nonmydriatic fundus camera with a snapshot FOV up to 67° external angle, which corresponds to a 101° eye angle. The wide-field fundus camera consists of a near-infrared light source (LS) for retinal guidance and a white LS for color retinal imaging. By incorporating digital image registration and glare elimination methods, a dual-image acquisition approach was used to achieve reflection artifact-free fundus photography.