Abstract
Near infrared emission spectroscopy (NIRES) allows the determination of the induction time (IT) of edible oils in accelerated oxidation experiments by monitoring the emissivity of a band at 2900nm, which corresponds to the formation of hydroperoxides. In this work, a new near infrared emission photometer dedicated to the determination of oxidative stability is described. The photometer presents several advantages compared to the previously reported NIRES instrument, such as lower cost and extreme simplicity of design and maintenance. The results obtained in the evaluation of the proposed instrument were compared with the official Rancimat method and instrument. The significant advantages include: faster analysis, lower sample consumption and operational simplicity. It is demonstrated that the procedure for determination of oxidative stability of oils can be significantly simplified and performed by measuring the sample emission at only one spectral region centered at 2900nm. Also, the proposed instrument and method present precision equivalent to the Rancimat method (coefficient of variation=5.0%). A significant correlation between the methods has been found (R2=0.81).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.