Abstract

An instrument has been built to measure the mode field profile of optical waveguides by the near field method. It incorporates a confocal microscope that allows accurate focusing of the waveguide end. The instrument has been used to measure the effect of defocus on measured fiber mode field diameters (MFD's). The amount of defocus must be less than a few micrometers if errors in the measured MFD are to be kept less than 0.1 /spl mu/m. Other focusing criteria can give errors of 2% or more. >

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.