Abstract

BackgroundIn order to efficiently perform laparoscopic microwave ablation of liver tumours precise positioning of the ablation probe is mandatory. This study evaluates the precision and ablation accuracy using the innovative laparoscopic stereotactic navigation system CAS-One-SPOT in comparison to 2d ultrasound guided laparoscopic ablation procedures.MethodsIn a pig liver ablation model four surgeons, experienced (n = 2) and inexperienced (n = 2) in laparoscopic ablation procedures, were randomized for 2d ultrasound guided laparoscopic or stereotactic navigated laparoscopic ablation procedures. Each surgeon performed a total of 20 ablations. Total attempts of needle placements, time from tumor localization till beginning of ablation and ablation accuracy were analyzed.ResultsThe use of the laparoscopic stereotactic navigation system led to a significant reduction in total attempts of needle placement. The experienced group of surgeons reduced the mean number of attempts from 2.75 ± 2.291 in the 2d ultrasound guided ablation group to 1.45 ± 1.191 (p = 0.0302) attempts in the stereotactic navigation group. Comparable results could be observed in the inexperienced group with a reduction of 2.5 ± 1.50 to 1.15 ± 0.489 (p = 0.0005). This was accompanied by a significant time saving from 101.3 ± 112.1 s to 48.75 ± 27.76 s (p = 0.0491) in the experienced and 165.5 ± 98.9 s to 66.75 ± 21.96 s (p < 0.0001) in the inexperienced surgeon group. The accuracy of the ablation process was hereby not impaired as postinterventional sectioning of the ablation zone revealed.ConclusionThe use of a stereotactic navigation system for laparoscopic microwave ablation procedures of liver tumors significantly reduces the attempts and time of predicted correct needle placement for novices and experienced surgeons without impairing the accuracy of the ablation procedure.

Highlights

  • In order to efficiently perform laparoscopic microwave ablation of liver tumours precise positioning of the ablation probe is mandatory

  • With the SPOT system, targeting and ablation needle placement is possible from any angle

  • With the use of the stereotactic navigation system the mean number of attempts could be reduced to 1.15 ± 0.48 attempts (p = 0.0005)

Read more

Summary

Introduction

In order to efficiently perform laparoscopic microwave ablation of liver tumours precise positioning of the ablation probe is mandatory. Results The use of the laparoscopic stereotactic navigation system led to a significant reduction in total attempts of needle placement. Comparable results could be observed in the inexperienced group with a reduction of 2.5 ± 1.50 to 1.15 ± 0.489 (p = 0.0005) This was accompanied by a significant time saving from 101.3 ± 112.1 s to 48.75 ± 27.76 s (p = 0.0491) in the experienced and 165.5 ± 98.9 s to 66.75 ± 21.96 s (p < 0.0001) in the inexperienced surgeon group. Conclusion The use of a stereotactic navigation system for laparoscopic microwave ablation procedures of liver tumors significantly reduces the attempts and time of predicted correct needle placement for novices and experienced surgeons without impairing the accuracy of the ablation procedure

Objectives
Methods
Results
Conclusion