Abstract

Using a digital simulation method, we analyzed the relationship between natural frequency (fn) and damping coefficient (Zeta) of the catheter-manometer system required for high-fidelity measurement of the pulmonary arterial pressure. The pulmonary artery pressure waveform was obtained with a catheter-tip transducer and it was fed into a dynamic simulator programmed on a computer. The original waveform and the output of the simulator were compared and judged visually for the fidelity. From this analysis, the combination of fn and Zeta was obtained and was plotted on a fn-Zeta diagram. It showed as an area, which was convex on the left side and open on the right side. The left-convex endpoint was located at a damping coefficient of about 0.7. At a lower heart rate, this area was extended to the lower frequency side, while, at a higher heart rate, this area was limited to the higher frequency side. The fn-Zeta diagram was also constructed theoretically by calculating the relations between natural frequencies and damping coefficients of a second order system with the amplitude and phase error tolerance set at +/-5% respectively.