Abstract

This study proposes the use of the inversion recovery T1-weighted turbo field echo (IR-T1TFE) sequence for myocardial T1 mapping and compares the results obtained with those of the modified Look-Locker inversion recovery (MOLLI) method for accuracy, precision, and reproducibility. A phantom containing seven vials with different T1 values was imaged, thereby comparing the T1 measurements between the inversion recovery spin-echo (IR-SE) technique, MOLLI, and the IR-T1TFE. The accuracy, precision, and reproducibility of the T1-mapping sequences were analyzed in a phantom study. Fifteen healthy subjects were recruited for the in vivo comparison of native myocardial T1 mapping using MOLLI and IR-T1TFE sequences. After myocardium segmentation, the T1 value of the entire myocardium was calculated. In the phantom study, excellent accuracy was achieved using IR-T1TFE for all T1 ranges. MOLLI displayed lower accuracy than IR-T1TFE (p =0.016), substantially underestimating T1 at large T1 values (> 1000ms). In the in vivo study, the first mean myocardial T1 values ± SD using MOLLI and IR-T1TFE were 1306 ± 70ms and 1484 ± 28ms, respectively, and the second were 1297 ± 68ms and 1474 ± 43ms, respectively. The native myocardial T1 obtained with MOLLI was lower than that of IR-T1TFE (p < 0.001). The reproducibility of native myocardial T1 mapping within the same sequence was not statistically significant (p = 0.11). This study demonstrates the utility and validity of myocardial T1 mapping using IR-T1TFE, which is a common sequence. This method was found to have high accuracy and reproducibility.