Abstract
The narrow gap of HbA1 value of mass fraction between "normal" (< 6.0%) and "diabetes" (≥ 6.5%) necessitates tight control of inter-assay standardization, assay precision, and trueness. This survey was initiated to obtain knowledge of the current situation of internal quality control (IQC) practice for HbA(1c) in China and find out the most appropriate quality specifications. Data of IQC for HbA(1c) in 331 institutions participating in the national proficiency testing (PT) programs in China were evaluated using four levels of quality specifications, and the percentages of laboratories meeting the quality requirement were calculated to find out the most appropriate quality specifications for control materials of HbA(1c) in China. The IQC data varied vastly among 331 clinical laboratories in China. The measurement of control materials covered a wide range from 4.52% to 12.24% (inter-quartile range) and there were significant differences among the CVs of different methods, including LPLC, CE-HPLC, AC-HPLC, immunoturbidimetry, and others. Among the four main methods, CE-HPLC and AC-HPLC achieved a better precision. As we can see, the performance of laboratories for HbA(1c) has yet to be improved. Clinical laboratories in China should improve their performance with a stricter imprecision criteria.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.