Abstract

BackgroundSeveral types of highly-crosslinked polyethylene with different manufacturing processes and mechanical properties are commercially available, including HIGHCROSS (Medacta). The aim of this registry study is to ascertain the long-term safety of the HIGHCROSS liners in total hip arthroplasty and compare the revision rates with its contemporaries using real-world data to establish their safety and generalizability when used by multiple surgeons. MethodsThe Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) and the Swiss National Joint Registry (SIRIS) were selected for the large number of users of Medacta implants and the availability of long-term results. Three reports from SIRIS and 4 reports from AOANJRR were examined to establish the overall and Kaplan-Meier (KM) cumulative revision rate. ResultsAccording to SIRIS, the HIGHCROSS liner was the most commonly used bearing surface with Medacta stems with overall revision rates for wear/osteolysis at 9.7 years of 0.04% and 0.03%, respectively, for AMIStem and Quadra. Based on AOANJRR, the KM cumulative revision rate for any reason of Medacta stems with HIGHCROSS liners was lower than that with ceramic liners at 3 years for MasterLoc and at 10 years for Quadra. The KM cumulative revision rate for any reasons of Medacta cementless cups with HIGHCROSS liners at 12 years was lower than the comparator made of all other implants and tribological couplings (1.6% vs 2.1%). ConclusionsThis real-world data proves that long-term HIGHCROSS survival rates are comparable to other modern bearing surfaces.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call