Abstract
Since 1998, the National Laboratory for Ionizing Radiation Metrology has lead an intercomparison program for activity measurements of radiopharmaceuticals administered to patients in the Nuclear Medicine Services (NMS) with the purpose of promoting the quality control. In this work, intercomparison results with the radionuclides 67Ga, 123I, 99mTc and 201Tl from the beginning of the program to the present will be presented establishing as a consequence, the radionuclide calibrators’ traceability used by the participants. The analysis of the results showed that 68% were within the limits of ±10% required by the Brazilian Regulatory Authority that demands this accuracy band for the radiopharmaceutical measurements in the NMS. The results also demonstrate an improvement in the measurement quality that this program has provided since its beginning.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.