Abstract
Air kerma-area product (PKA), cumulative air kerma at patient entrance reference point, fluoroscopy time and number of images were retrospectively collected from 15 hospitals in Lebanon for 11282 fluoroscopically-guided interventional (FGI) procedures between March 2016 and November 2018. National diagnostic reference levels (NDRLs) were established based on the third quartile of the distribution of median values of exposure parameters per department for 27 types of FGI procedures. NDRLs were in line with international DRLs except for coronary angiography (CA), percutaneous coronary interventions (PCI) and transcatheter aortic valve implantation (TAVI) which require optimisation. Additionally, following the National Council on Radiation Protection and Measurements report 168, PCI, TAVI, triple chamber pacemaker implantation, endovascular aortic repair, nephrostomy, kyphoplasty and percutaneous transhepatic biliary drainage were classified as potentially high-dose procedures with >5% of the patients with PKA exceeding 300 Gycm2. The established NDRLs will promote dose optimisation and patient radiation protection.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.