Abstract
To compare the feasibility of continuous positive airway pressure (CPAP) support generated by high flow nasal cannula with conventional CPAP for prevention of reintubation among preterm infants with a birth weight of <or=1,250 g. Preterm infants were randomized to CPAP generated via high flow cannula or the Infant Flow Nasal CPAP System (VIASYS, Conshohocken, PA, USA) at extubation. Primary outcome was incidence of reintubation within 7 days. Secondary outcomes included change in oxygen use and frequency of apnea and bradycardias postextubation. Forty neonates were randomized. Twelve of 20 infants randomized to high flow cannula CPAP were reintubated compared to three of 20 using Infant Flow (P=0.003). The high flow cannula group had increased oxygen use and more apneas and bradycardias postextubation. CPAP delivered by high flow nasal cannula failed to maintain extubation status among preterm infants <or=1,250 g as effectively as Infant Flow CPAP.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.