Objective parameters to assess the physical flow conditions of breathing are scarce and decisions for surgery, e.g. nasal septum correction, mainly rely on subjective surgeon judgment. To define decision supporting parameters, we compare laser Doppler anemometry (LDA) and numerical computational fluid dynamic simulations (CFD) of the airflow velocity vector fields in the nasal cavity, including lattice Boltzmann (LB) and finite volume methods (FVM). The simulations are based on an anonymous patient CT dataset with septal deviation. LDA measurements are preformed using a 3D printed model. Nasal airflow geometry is randomly deformed in order to approximate surgical changes. The root-mean-square velocity error near the nasal valve of laser Doppler anemometry and lattice Boltzmann simulations is 0.071. Changes in geometry similarly affect both measurement and simulation.