Abstract

A recent report indicated significant interference of naproxen with total bilirubin measurement when using the Jendrassik and Grof method. We explored the possibility of avoiding this interference by using a different method. We observed that naproxen has no effect on total bilirubin measurement by any method, but O-desmethylnaproxen (the naproxen metabolite) interfered significantly with total bilirubin assays using the Jendrassik and Grof method (Beckman Coulter [Brea, CA] and Siemens [Tarrytown, NY] bilirubin method). The Roche (Indianapolis, IN) bilirubin assay based on a different method was not affected. When serum samples of 9 patients receiving naproxen were analyzed for total bilirubin level, we observed good correlation in total bilirubin values measured by all 3 methods in 7 of 9 patients with therapeutic naproxen concentrations. In contrast, in 2 patients with elevated naproxen levels, the total bilirubin values observed by using the Beckman and Siemens methods were significantly higher than the value obtained by the Roche assay. The naproxen metabolite interferes with assays based on the Jendrassik and Grof method but not with the Roche bilirubin assay.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.