Abstract

Limiting contamination of liquid chromatography/tandem mass spectrometry (LC-MS) systems and reducing the downtime associated with maintenance and cleaning is essential for productivity. We developed a simple device that creates a gas curtain barrier to prevent ions from entering the MS inlet. The gas can be quickly and easily applied when certain contaminant ions are known to elute. We show that the device can prevent the buildup of contaminants on the heated transfer capillary following >100 injections of a crude tissue lysate and improves peptide identifications. The device may provide a promising approach toward improving instrument robustness.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.