Abstract

The relationship between dust size and penetration for a static horizontal elutriator (Sibata C-30) was measured in calm air. The elutriator as a low-volume air sampler is widely used as a dust size classifier in work environment measurements. The actual penetrations were compared with the theoretical models of the sampler and with sampling convention for respirable dust in work environment measurement. The sampling convention was recently introduced into the Japanese standard for work environment measurement and is based on the ISO 7708 respirable dust convention. The bias of sampled masses from the respirable dust was calculated for two flow rates of the sampler, i.e., 50% cut sizes of 4 microm and 5 microm, from measured penetration curves. The bias of the sampler was overestimated in the 5 microm, 50% cut condition and underestimated in the 4 microm, 50% cut condition for most workplace sampling situations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.