The performance of specialy designed an angiographic table for examination of legs and its image quality are discussed. A rotating long cassette changer having 6 cassettes with 10 X 36 in is placed under the table. The table and long cassette changer are tilted through a range from the vertical to horizontal position. The distance from the focus to the table top is 193.7cm and the focus-film distance is 200cm. The image quality are evaluated by the MTFs. For a vibration caused by rotating the long cassette changer, the image quality are slightly degraded. However, it is not the main reason of the degradation of overall image quality.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.