Abstract

It has been noted that the manual settings of region of interest (ROI) to the single-photon-emission-computed-tomography (SPECT) slice lacked objectivity when the fixed quantity value of regional cerebral blood flow (rCBF) was measured previously. Therefore, we jointly developed software "Brain ROI" with Daiichi Radioisotope Laboratories, Ltd. (Present name: FUJIFILM RI Pharma Co., Ltd.) The software normalized an individual brain to a standard brain template by using Statistical Parametric Mapping 2 (SPM 2) of the easy Z-score Imaging System ver. 3.0 (eZIS Ver. 3.0), and the ROI template was set to a specific slice. In this study, we evaluated the accuracy of this software with an ROI template that we made of useful size and shape, in some clinical samples. The method of automatic setting of ROI was the objective. However, we felt that we should use the shape of the ROI template without the influence of brain atrophy. Moreover, we should see normalization of the individual brain and confirm the accuracy of normalization. When normalization failed, we should partially correct the ROI or set everything by manual operation for the operator. However, it was thought that this software was useful if the tendency was understood because examples of failure were few.