Abstract

A thermal diffusivity measuring system employing a closed cycle helium refrigerator was developed, which also enabled the thermal conductivity measurement under an identical experimental setup. The diffusivity was measured by a discretional heating method and the conductivity was measured by a steady-state heat flow method. The errors and the accuracy in measurements were evaluated by measuring the diffusivities of an austenitic stainless steel standard sample (SRM 1460), a Pyrex glass (Corning #7740) and a high purity copper between 12 to 200K. The diffusivity ranging from 10-3 to 2cm2/s could be determined with an uncertainty of 3%. The results agreed with the reported values within 10%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.