Abstract

Based on spectral interference technique, a method for simultaneously measuring the thickness and refractive index of a glass was proposed. This method used the Michelson optical path to solve the interference signal received by the spectrometer through a Fourier transform algorithm to obtain the modulation period of the spectral interference fringe. The modulation period was determined according to the optical path difference formed between the measuring arm and the reference arm before and after the glass sample to be tested was placed in the measuring arm, and the geometric thickness and refractive index of the glass sample can be obtained. The method did not need mechanical scanning delay line and adopted the improved phase extraction algorithm in the Fourier domain, which improved the anti-interference ability of the measurement system and the detection speed was fast. The experimental results show that the thickness measurement accuracy of the glass sample is better than±1μm, and the refractive index measurement accuracy is ±5×10-4.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.