Abstract

Mechanical accuracy of a stereotactic irradiation system using a micro multi-leaf collimator (mMLC), Elekta DMLC, has been evaluated. Measurements were made to obtain transmission, leakage, penumbra, and positioning accuracy of the DMLC leaf for a 6 MV photon beam. Mechanical accuracy and long term stability of a linac isocenter was also evaluated. The resulting transmission, along a line perpendicular to the leaf movement, was 0.31±0.01%, and the leakage from the closed opposing leaf pairs was 0.39±0.01%. The measured penumbra, at a depth incurring maximum dose, was 2.37±0.16 mm toward the leaf end and 2.14±0.18 mm toward the leaf side for various field sizes. The leaf gap width error, of 0.10±0.08 mm, was obtained by analyzing picket fence test results. The maximum leaf positioning error, of 0.14±0.06 mm, was obtained by analyzing the log file for a various gantry angles during an arc delivery. The isocenter accuracy was within a radius of 1 mm, without any recalibration for two years. In conclusion, our stereotactic irradiation system using DMLC was capable of providing accurate stereotactic treatment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.