Abstract

A humidity generator using divided flow method was constructed for trial. The principle, the structure and the characteristics of the generator are described. The accuracy of mass flow meters and the efficiency of saturators were investigated. The cause of errors of the generator is discussed and the accuracy of the generator is estimated.The generator was compared with the high precision two-pressure and two-temperature humidity generator, using a quartz crystal dew-point hygrometer. The humidity value of the air produced by these two generators showed a good agreement with an accuracy of ±0.5% and it is confirmed that a divided flow humidity generator can be used as a working standard for the calibration of humidity sensors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.