Abstract
Variation in the luminance ratio of a cathode ray tube(CRT)monitor and the ultrasonographic images at different levels of ambient light(0-150 lux)was investigated to obtain optimum ambient light in the ultrasonography suite. The maximum and minimum luminances of test patterns and ultrasonographic images were measured after three technicians independently optimized the brightness and contrast of the CRT monitor and ultrasonographic images at different levels of ambient light. Furthermore, the luminance ratio was calculated from the maximum luminance divided by the minimum luminance. When ambient light increased, it was difficult for the technicians to optimize the brightness and contrast settings of the CRT monitor to maintain a high luminance ratio at 0 lux. The luminance ratio decreased rapidly as ambient light increased up to 20 lux. However, the luminance ratio decreased gradually when ambient light was higher than 20 lux. It is necessary to take into consideration the ambient light to maintain a high luminance ratio of ultrasonographic images.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.