Abstract

SUMMARY We have developed a novel reconstruction method for a laser interferometer to determine the radial electron density profile in a cylindrical plasma vessel under an axial symmetrical assumption with spline interpolation. This method can extend the reconstructible area out of the line-integrated density measurement area by using the additional local measurement data. Although both the ordinary Abel inversion and the basis function expansion method cannot reconstruct well out of the measurement area, the new method can reconstruct distributions even with nonzero electron density on the wall. In order to confirm the validity of the method, the error sensitivity is evaluated by Monte Carlo simulation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.