Abstract

We measured the effective dose received by the person assisting the patient at diagnostic X-ray examination. Measurement was done when a patient's chest, abdomen, lumbar vertebrae, hip joint, skull, cervical vertebrae, or knee joint was examined by radiography. A body phantom including human bones exposed to radiation was used in the role of the patient. Some exposure conditions for these measurements were the same as those used routinely in computed radiography. Effective dose was measured directly with an ionization survey meter. As a result, the effective dose of the person assisting with axial projection of the hip joint was 124 microSv, which was higher than that for other regions and projections. If the assisting person helped a patient without using any protective device, the effective dose would be low enough to ignore. However, because medical staff are frequently exposed to radiation, optimal protection is crucial to prevent unnecessary radiation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.