Abstract

In order to avoid postoperative liver failure after hepatectomy, the evaluation with liver functiontests and the resecting volume should be done actually.By examining our clinical cases retrospectively and prospectively, we got the formula to calculatethe safe resecting volume of the liver tissue especially in the cases of hepatectomy using the Microwave Tissue Coagulator (Microtaze®). With the indicators of the serum disappearance rateof indocyanine green (ICGK), the percent activity of the hepaplastintest (HPT), the volume of the tumor (VT) calculated from the diameter of the main tumor as a sphere, and the long diameter (R) of the oval shaped resecting line which mapped out on the surface of the liver, the formula “460×ICGK+VT (cm3)” was clinically used in the cases of the typical hepatectomy, and the formula “500×ICGK+3.7×HPT+VT-7.5×R-100 (cm3)” was used in the cases of atypical hepatectomyas the safe resecting liver volume.They will be useful in the cases of hepatectomy using the Microwave Tissue Coagulator.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.