Abstract
To investigate the accuracy and usefulness of combined transcutaneous carbon dioxide (PtcCO2) and oxygen (PtcO2) measurement compared with arterial blood gas analysis (PaCO2 and PaO2), 10 patients undergoing laryngomicrosurgery under total intravenous anesthesia were studied. The PtcCO2 values revealed a high degree of correlation with PaCO2 during anesthesia (r=0.96) and after anesthesia (r=0.93). The PaO2 values revealed a generally good correlation with PaO2 during anesthesia (r=0.81), but not after anesthesia (r=0.23). Capnography values are invalid after anesthesia, and an arterial catheter is not always indicated and feasible in microlaryngoscopy. Transcutaneous devices provide an effective method for the non-invasive monitoring of PCO2 during and after anesthesia, in situations whereas continuous, precise control of CO2 levels is desired such as laryngomicrosurgery.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.