A method of evaluating the performance of computed tomography (CT) with equivalent resolution images was investigated. Generally, in performance evaluations of CT, the resolution property is measured by the wire method, and the noise property is measured from noise images of a cylindrical water phantom. The signal-to-noise ratio (SNR) is then calculated for the integrated evaluation. Our proposed method enabled perceptual integrated evaluation by using equivalent resolution images created with frequency processing. The frequency-processing factor was calculated from the ratio of the modulation transfer factors of two models of CT, and the image of one of the two was processed by the factor. Because these processed images have resolution equivalent to images of the other CT, the perceptual evaluation with noise images becomes effective. In this investigation, images of a water phantom and a middle contrast resolution phantom were employed. Perceptual comparison of the amount of noise with equivalent resolution images could be performed easily, and effective performance evaluation was achieved. Therefore, our proposed method is useful for noise property and performance evaluation of CT.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.