Abstract
In a sound velocity measurement in solid, the properties of electrical source and ultrasonic transducer bonded to a solid material become the error sources. In this paper, the error sources are estimated from the simulation using transmission line model of a measurement system and their experiments. A novel method for correcting the measured values of sound velocity by a numerical fitting is suggested. The usefulness of our method is demonstrated by comparing with other methods. Sound velocity measurement with the accuracy better than 1% without regard to the electrical and mechanical effects or bonding layers effects have been successfully accomplished.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Society of Materials Engineering for Resources of Japan
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.