Abstract

A novel concept for an optical multilayer ultrasonic hydrophone with the sensing film deposited on a triangular pyramid glass substrate is proposed. Using the calculation model for the spectral coefficients' derivatives of a dielectric multilayer optical coating, the acousto-optic sensitivity characteristic of the hydrophone is analyzed with different measurement laser polarizations and incident angles. We present a reasonable method and adjusting strategy for the optimum working point selection of the ultrasound measurement. Analytic results show that the novel hydrophone possesses all the other merits of a plate glass substrate optical multilayer hydrophone but with improved detection sensitivity. A longer measurement time without distortion decreases the difficulty of high frequency signal circuits. Spatial split of the ultrasound signal caused by the substrate's triangular pyramid roof simplifies the spatial spot area correction, which contributes to the accurate calibration of the hydrophone's wideband frequency response.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.