Abstract

It is empirically said that the application of the compensating filter leads to a decrease in the exposure dose of the filter-inserted area and an increase in that of the otherwise area. Using the area-dosimeter, comparison was made of exposure doses by the application of the above filter and the otherwise filter in head simple X-P, abdominal angiography and lower extremity X-P. Using the filter for head simple X-P and Mix-Dp phantom, measurement was made of the absorbed dose at the 5 cm-depth to compare the rate of decrease in absorbed dose between the above both areas. Head simple X-P gained a decrease in area dose of 29%. The absorbed dose at the 5 cm-depth in the phantom experiment showed a decrease of over 26% at the filter-inserted area, but little increase at the otherwise area. The above results indicated the interposition of the filter between the X-ray tube and the object to lead to decreases not only in the area dose but also in the patient's exposure dose.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.